Overdue Calibration Finder

Calibration interval: what the standards ask, and what they leave to you

None of the standards here gives an interval for any instrument. They ask for calibration or verification at set intervals or before use, and for the organization to set, record and review those intervals. The finder reads the interval in your register and adds it to the last calibration date; it never sets or suggests one.

What the standards say

Named, not quoted

ILAC-G24 / OIML D10 give guidance on setting and reviewing calibration intervals from the history of each instrument (as-found results, drift, use). ISO 10012 sets out a measurement management system. Both are named here and not stated.

How the finder reads an interval

An interval reads as months (12 months, 12M, 12 mo, 1 yr, annual, 2 years, biennial, 6 mo, semi-annual, quarterly, monthly), weeks (4 weeks, 2 wk, weekly) or days (365 d, 90 days, daily). Calendar months are added as months: 31 January plus one month is the last day of February. "Before use" is read as verification before each use, with no due date to work out. Anything else is shown as not read, never guessed.

PastedRead asLast calibration 31/01/2026 gives
12 months, 12M, 1 yr, annual12 months31/01/2027
6 mo, 6M, semi-annual6 months31/07/2026
quarterly3 months30/04/2026
1 month, monthly1 month28/02/2026
90 days90 days01/05/2026
2 years, biennial24 months31/01/2028
before each useno due datenone
biannualnot read: twice a year or every two years?none

Where an interval shows up in the findings

AS9100D 7.1.5Monitoring and measuring resources

The resources needed for monitoring and measurement to give valid, reliable results are identified and provided; they are suitable for the task and kept in good order, and records show they are fit for purpose. 9100 adds a process for calling in monitoring and measuring equipment that needs calibration or verification, and a register of such equipment showing its type, unique identity, location, and the method, frequency and acceptance criteria for calibration or verification; the register covers test hardware and test software, automated test equipment, plotters that produce verification data, and equipment owned by employees or supplied by customers. Where traceability of measurement is required, or the organization regards it as essential, equipment is calibrated or verified at set intervals against standards traceable to international or national standards (with the basis recorded where none exists), marked to show its status, and protected against adjustment, damage or deterioration. Under 9100, calibration and verification take place in suitable environmental conditions. If equipment is found to be unfit, the organization establishes whether earlier results were compromised and acts accordingly.

What a certification body auditor asks to see: Register of monitoring and measuring equipment with the six 9100 fields; Recall process for equipment due for calibration; Traceable calibration records, status identification and impact assessments of out-of-tolerance findings
Where registers usually fall short: Test software and ATE outside the calibration register; Personally owned gauges uncontrolled; No recall process, so equipment runs past its due date
Source: AS9100D, aerospace quality management