Overdue Calibration Finder
Findings · finding 3 of 12

No due date can be worked out: no interval or no last calibration recorded

Without a last calibration date and an interval the register cannot show when the instrument is due back, and so cannot show that it is in date. AS9100D asks for the method and frequency in the register itself; ISO/IEC 17025 asks for the next due date in the equipment record.

When the finder raises it

Fires for an instrument in service with no last calibration date or no interval the finder can read. "Before each use" is an interval with no due date and does not fire. Where the register has its own due date, it is used for the schedule and the finding still names the missing column.

A question for the calibration lead

What is the interval, and when was it last done?

Clauses behind it

5 across the standards
StandardClauseApplies to
ISO 9001ISO 9001 7.1.5 Monitoring and measuring resourcesa manufacturer holding ISO 9001 certification
AS9100DAS9100D 7.1.5 Monitoring and measuring resourcesan aerospace supplier holding AS9100 certification
ISO/IEC 17025ISO/IEC 17025 6.4 Equipmenta testing or calibration laboratory accredited to ISO/IEC 17025
ISO 13485ISO 13485 7.6 Control of monitoring and measuring equipmenta medical device maker holding ISO 13485 certification
IATF 16949IATF 16949 7.1.5.2.1 Calibration and verification recordsan automotive supplier holding IATF 16949 certification
ISO 9001 7.1.5Monitoring and measuring resources

Where monitoring or measurement is used to check that products and services conform, the organization identifies and provides the resources needed for results to be valid and reliable. It makes sure those resources suit the kind of monitoring and measurement done and are maintained so they stay fit for purpose, and it retains documented information as evidence that they are. Where traceability of measurement is required, or the organization judges it essential to confidence in the results, measuring equipment is: calibrated or verified, or both, at set intervals or before use, against standards traceable to international or national ones (and where none exist, the basis used is recorded); identified so its status is known; and protected from adjustment, damage or deterioration that would make its calibration status or later results invalid. If equipment proves unfit for its purpose, the organization judges whether earlier results remain valid and acts as needed.

What a certification body auditor asks to see: Register of monitoring and measuring equipment with calibration status; Calibration certificates traceable to national standards; Procedure for handling out-of-tolerance equipment and assessing previous results; Identification and safeguarding of equipment
Where registers usually fall short: Equipment used for acceptance not in the calibration register; Out-of-tolerance findings not followed by an assessment of prior results; Software used for measurement not validated
Source: ISO 9001:2015, quality management systems

Two questions on the row, never counted as findings

See it on the specimenAll twelve findings