Overdue Calibration Finder
Mass and force · instrument class

Weighing scale

A weighing scale is one of 8 classes the finder reads in the mass and force group. Scales, balances, test weights, torque tools, force gauges and load cells: where an as-found reading out of tolerance usually means product was weighed, torqued or loaded against the wrong figure.

Calibrating a weighing scale

vendor-neutral, no interval given
What a calibration usually checks
Readings with test weights across the range, eccentric loading, repeatability.
Common as-found problems
Load cell drift, damage from shock loads, debris under the platform.
Calibration, and the routine check
A calibration compares it with a reference traceable to national or international standards and records the as-found result; the routine check (a check weight before use) confirms it is working on the day and does not replace the calibration. Your procedure sets both, and the interval.
What the register row should carry
The asset ID on the instrument and its label, the last calibration date and the certificate reference, who did it (in house against which reference standard, or which external provider), the as-found result and, if it was out of tolerance, the impact assessment reference, and the due date worked out from your interval.

What the calibration register should record for it

ColumnWhat goes in it
Asset IDthe number on the instrument and its label, unique in the register
Type and descriptionwhat it is and its range, in the words your register uses
Location and ownerwhere it lives and who answers for it, so the recall list can find it
Control typeoptional: calibration, verification, intermediate check, qualification or mapping; the wording of the due date follows it
Last calibrationthe date of the last calibration or verification
Intervalthe interval your procedure sets, in months, weeks or days, or before each use
Due dateoptional: the finder works it out and shows where yours disagrees
Cal byin house or external
Lab accredited to ISO/IEC 17025yes, no or unknown, for an external lab
Certificate numberthe index to the calibration record
Traceable to a national standardyes or no
As foundin tolerance or out of tolerance at the last calibration
Used for product acceptanceyes or no
Statusin service, out of service, reference only, lost, quarantined or retired
Impact assessmentyes, no, or a reference, for any out-of-tolerance reading

The due date

worked out, never set

The due date is the last calibration plus the interval in your register, shown with its arithmetic. The interval is yours to set and review; the finder never suggests one, and no typical interval enters any arithmetic on this site.

A line that places here

example

A-015 | platform scale 300 kg | 12/03/2025 | 12 months | Inspection | yes | in tolerance

See the specimen register run

What the finder reads on these lines

12 findings

Clauses, by standard

10 cited
StandardClauseApplies to
ISO 9001ISO 9001 7.1.5 Monitoring and measuring resourcesa manufacturer holding ISO 9001 certification
AS9100DAS9100D 7.1.5 Monitoring and measuring resourcesan aerospace supplier holding AS9100 certification
ISO 13485ISO 13485 7.6 Control of monitoring and measuring equipmenta medical device maker holding ISO 13485 certification
ISO/IEC 17025ISO/IEC 17025 6.4 Equipmenta testing or calibration laboratory accredited to ISO/IEC 17025
ISO/IEC 17025ISO/IEC 17025 6.5 Metrological traceabilitya testing or calibration laboratory accredited to ISO/IEC 17025
IATF 16949IATF 16949 7.1.5.2.1 Calibration and verification recordsan automotive supplier holding IATF 16949 certification
21 CFR 21121 CFR 211.68 Automatic, Mechanical, and Electronic Equipmenta drug maker under 21 CFR 211
21 CFR 5821 CFR 58.63 Maintenance and Calibration of Equipmenta laboratory running nonclinical studies under 21 CFR 58
IATF 16949IATF 16949 7.1.5.1.1 Measurement systems analysisan automotive supplier holding IATF 16949 certification
21 CFR 21121 CFR 211.160 Laboratory controls: general requirementsa drug maker under 21 CFR 211

The calibration clause, set out

ISO 9001 7.1.5Monitoring and measuring resources

Where monitoring or measurement is used to check that products and services conform, the organization identifies and provides the resources needed for results to be valid and reliable. It makes sure those resources suit the kind of monitoring and measurement done and are maintained so they stay fit for purpose, and it retains documented information as evidence that they are. Where traceability of measurement is required, or the organization judges it essential to confidence in the results, measuring equipment is: calibrated or verified, or both, at set intervals or before use, against standards traceable to international or national ones (and where none exist, the basis used is recorded); identified so its status is known; and protected from adjustment, damage or deterioration that would make its calibration status or later results invalid. If equipment proves unfit for its purpose, the organization judges whether earlier results remain valid and acts as needed.

What a certification body auditor asks to see: Register of monitoring and measuring equipment with calibration status; Calibration certificates traceable to national standards; Procedure for handling out-of-tolerance equipment and assessing previous results; Identification and safeguarding of equipment
Where registers usually fall short: Equipment used for acceptance not in the calibration register; Out-of-tolerance findings not followed by an assessment of prior results; Software used for measurement not validated
Source: ISO 9001:2015, quality management systems